FTB3001,FTB3002
Positive Displacement (PD) Flowmeters

FTB3001,FTB3002 : Positive Displacement (PD) Flowmeters
Kliknìte zde pro vìtší obrázek.

90,900.00  Kč *
     FTB3002

  Toothless Oval Gear Construction (Patented)
  High Accuracy
  Compact Design
  High Resolution Electronic Output
  Temperatures to 176°C (350°F)
  Pressure Rated to 3000 PSIG with 1 NPT Fitting Standard
  Multiple Bearing Options
  316 Stainless Steel Construction (Standard)
  No Special Gears Needed for High Viscosity
  Linearity Unaffected by Viscosity Changes
  1 MNPT End Fitting Standard, Flange Optional

The oval-shaped smooth gears in the FTB3000 Series are used to displace a precise volume of fluid,which is passed through th emeasurement chamber during each revolution. The toothless design,working with the fluid being measured, provides a complete viscous seal within the measuring chamber. This sealing effect provides for greater flow measurement accuracy.

The displacement of the fluid is translated via a magnetic or Hall-effect pick-up into a pulse output that is proportional to the flow.

The unique design of the FTB3000 oval gear meter incorporates two smooth(toothless) oval gears positioned 90 degrees out of phase. The measurement gears are held together by two timing gears. The standard meter contains no seal between the measurement chamber and the timing chamber. The timing gears have a pitch diameter equal to the outside diameter of the measurement gears. The flow through the meter measurement chamber follows the path of least resistance. Therefore, no liquid passes through the center cavity between the measurement gears. The fluid is displaced from the inlet to the outlet via the area between the smooth oval gear and the inner diameter of the meter housing.

Conventional oval gears have teeth that mesh. One of the advantages of smooth oval gears is that the viscous flow does not get trapped or squeezed between the gears. Typically, oval gears need to have cuts made in the teeth to allow high viscosity fluids to pass. These cuts result in a decrease in accuracy.

SPECIFICATIONS
Service Fluid: Clean liquids, max particle size 3.175 mm (0.125")
Accuracy: ±0.25% of reading
Viscosity: 100 cSt or greater
Repeatability: ±0.05%
Operating Temperature:
FTB3001: -240 to 176°C (-400 to 350°F)
FTB3002: -40 to 150°C (-40 to 302°F)
Operating Pressure: 3000 psig standard
Wetted Parts: 316 SS body and gears with PEEK™ gear seats
Bearings: Shielded, self-lubricating 440 SS ball bearings
Connections: 1 MNPT
Pick Up Coil: Magnetic type or Hall Effect (6 to 24 Vdc power)
Calibration: Ten-point calibration traceable to NIST @ 100 centistokes
Weight: 7.53 kg (16.6 lb)


 Èíslo modelu  Cena (bez DPH)  Popis  Ks
 FTB3001    90,900.00 Kč   2 to 25 GPM rate, magnetic pick-up
 FTB3002    90,900.00 Kč   0.02 to 25 GPM rate, Hall-effect sensor
 U24Y175    5,160.00 Kč   24 Vdc, 175 mA power supply
 DPF701    8,580.00 Kč   Panel meter for rate or total
 MS3106F-10SL-3S    2,130.00 Kč   Replacement mating connector for FTB3000 Series
Obsah košíku Vyprázdnit košík Objednat
 * využijte možnost konzultovat ceny pøímo s našimi prodejci  (tel: 596 311 899 - prodejní oddìlení, e-mail: viz. kontaktní informace)
 Poznámka: Add suffix “-150LBF” (150 lb stainless steel flange) or “-300LBF” (300 lb stainless steel flange) to the part number and $250 to the price.

Comes complete with operator’s manual, mating connector, and 10-point NIST calibration.


 Pøíklad objednávky: (1) FTB3002 Hall effect output, max temp. 150°C (302°F) unit with operator’s manual and 10-point NIST calibration = 90,900.00 Kč

 Literatura
   Stáhnout kompletní specifikace produktu ve formátu PDF
 Manuály
   Stáhnout  FTB3000 SERIES - Postive Displacement Meters
 Souvísející odkazy
   DPF70 Series: Frequency Input Ratemeter/Totalizer
   DPF700 Series: 1/8 DIN 6-Digit Rate Meter/Totalizer
   CL535: Frequency Calibrator with Totalizer