PCL1200
Portable, High Accuracy Multifunction Calibrator

PCL1200 : Portable, High Accuracy Multifunction Calibrator
Kliknìte zde pro vìtší obrázek.

71,400.00  Kč *
     PCL1200

  Measures and Sources T/C (13 Types), RTDs (13 Types), Ohms, Current, Voltage, Frequency
  Sources Pulse Trains
  Isolated mA/V Readback Circuit for Complete Transmitter Calibration
  Pressure Module Communication Port Compatible with OMEGA
PCL-PM Series Pressure Modules

  Built-In 24 V Supply Can Drive 4 to 20 mA Loops Up to 1000 Ohms
  Direct Entry of Custom RTD Coefficients (R0, A, B, C)
  All Source Modes Can be Programmed with Dedicated Setpoints to Speed Calibration and Linearity Tests
  Highest Accuracy in Class—to 0.015% of Reading
  Meets CE Requirements and is Designed to IEC 1010 Safety Standards
  Supplied with Full Rubber Boot

The PCL1200 multifunction calibrator has a feature set unmatched by other high-accuracy handheld calibrators in its price range. It is portable, provides the functions and accuracy of a laboratory-grade instrument, and has everything needed for virtually any calibration task. It can also measure and source thermocouples, RTDs, current, voltage, and frequency, and it can source pulse trains. It has a communications port for pressure modules and an isolated mA/V readback circuit. Arrow keys, direct numeric keypad entry, and 3 software-driven function buttons combine with a large, backlit, menu-driven graphics display to provide a highly intuitive yet powerful interface. A built-in 250 Ohm resistor for HartTM compatibility, compatibility with smart transmitters and PLCs, full fuseless protection, and a serial communications port for control using ASCII commands are additional features that make the PCL1200 indispensable for calibration tasks. A protective rubber boot is included with each unit.

Measuring Pressure with the PCL1200 With a PCL-PMA pressure module adaptor, the PCL1200 will work with all OMEGA PCL-PM pressure modules. See Specs and "To Order" chart.

SPECIFICATIONS (23 ±5°C Unless Otherwise Noted)
Voltage Read and Source:
Source: 0 to 20 Vdc
    Read:
        Isolated: 0 to 30 Vdc
        Non-Isolated: 0 to 20 Vdc
Thermocouple mV:
    Read/Source: -10 to 75 mV
Current (mA):
Source: 0 to 24 mA
    Read:
        Isolated: 0 to 24 mA
        Non-Isolated: 0 to 24 mA
Frequency (1 to 20 V Selectable Amplitude):
    cpm Source/Read: 2 to 600 cpm
    Hz Source/Read: 1 to 1000 Hz
    kHz Source/Read: 1 to 10 kHz
Pulse (Source Only; 1 to 20 V Selectable Amplitude): 1 to 30,000 pulses; 2 cpm to 10 kHz
Ohms:
    Source: 5 to 4000 O
    Read: 0 to 4000 O
Thermocouple Read and Source:
    Type J: -200 to 1200°C (-328 to 2192oF)
    Type K: -200 to 1370°C (-328 to 2498oF)
    Type T: -200 to 400°C (-328 to 752oF)
    Type E: -200 to 950°C (-328 to 1742oF)
    Type R: -20 to 1750°C (-4 to 3182oF)
    Type S: -20 to 1750°C (-4 to 3182oF)
    Type B: 600 to 1800°C (1112 to 3272oF)
    Type C: 0 to 2316°C (32 to 4201oF)
    Type XK: -200 to 800°C (-328 to 1472oF)
    Type BP: 0 to 2500°C (32 to 4532oF)
    Type L: -200 to 900°C (-328 to 1652oF)
    Type U: -200 to 400°C (-328 to 752oF)
    Type N: -200 to 1300°C (-328 to 2372oF)
RTD Read and Source:
    Ni120 (672): -80 to 260°C (-112 to 500oF)
    Pt100 (385): -200 to 800°C (-328 to 1472oF)
    Pt100 (3926): -200 to 630°C (-328 to 1166oF)
    Pt100 (3916): -200 to 630°C (-328 to 1166oF)
    Pt200 (385): -200 to 630°C (-328 to 1166oF)
    Pt500 (385): -200 to 630°C (-328 to 1166oF)
    Pt1000 (385): -200 to 630°C (-328 to 1166oF)
    Cu10: -100 to 260°C (-148 to 500oF)
    YSI400: 15 to 50°C (59 to 122oF)
    Cu50: -180 to 200°C (-292 to 392oF)
    Cu100: -180 to 200°C (-292 to 392oF)
    Pt385-10: -200 to 800°C (-328 to 1472oF)
    Pt385-50: -200 to 800°C (-328 to 1472oF)


Environmental
Operating Temperature: -10 to 50°C (14 to 122oF)
Storage Temperature: -20 to 70°C (-4 to 158oF)
Stability: ±0.005% of reading/°C (±5°C outside of 23°C)
Power Requirement: 6 Vdc
Batteries: 4 "AA" alkaline (included) or optional rechargeable
Mechanical
Dimensions: 220.9 H x 106.6 W x 58.4 mm D (8.7 x 4.2 x 2.3")
Weight: 863 g (30.5 oz)
Optional Accessory: Carrying case (Model No. CC572)
Accuracy Voltage: ±0.015% of reading, ±2 mV
Thermocouple: mV ±0.02% of reading, ±10 µV
Thermocouple Errors (In °C; Add 0.2 for Cold-Junction Compensation Error)
RTD Read/Source:
    Type J: 0.2°C Ni120 (672) 0.2°C
    Type K: 0.3°C Pt100 (385) 0.2°C
    Type T: 0.2°C Pt100 (3926) 0.2°C
    Type E: 0.2°C Pt100 (3916) 0.2°C
    Type R: 1.2°C Pt200 (385) 0.8°C
    Type S: 1.2°C Pt500 (385) 0.4°C
    Type B: 1.2°C Pt1000 (385) 0.2°C
    Type C: 0.6°C Cu10 1.4°C
    Type XK: 0.2°C YSI400 0.1°C
    Type BP: 0.9°C Cu50 0.4°C
    Type L: 0.2°C Cu100 0.3°C
    Type U: 0.25°C Pt385-10 1.4°C
    Type N: 0.4°C Pt385-50 0.4°C
Read Source:
    Current (mA): ±0.015% of reading, ±2 µA ±0.015% of reading, ±2 µA
    cpm: ±0.05% of reading, ±1 LSD ±0.05% of reading
    Hz: ±0.05% of reading, ±1 LSD ±0.05% of reading
    kHz: ±0.05% of reading, ±1 LSD ±0.125% of reading
Ohms (Works with All Pulsed Transmitters Down to 5 ms):
400 Ω Range: ±0.025% of reading, ±0.05 Ω ±0.025% of reading, ±0.05 Ω
4000 Ω Range: ±0.025% of reading, ±0.5 Ω ±0.025% of reading, ±0.5 Ω


 Èíslo modelu  Cena (bez DPH)  Popis  Ks
 PCL1200    71,400.00 Kč   High-accuracy multifunction calibrator
 PCL-PMA    12,600.00 Kč   Pressure module adaptor (one required for pressure)
 PCL-PM030G    30,150.00 Kč   External pressure module, 0 to 30 psig range
 PCL-PM100G    30,150.00 Kč   External pressure module, 0 to 100 psig range
 PCL-PM300G    30,150.00 Kč   External pressure module, 0 to 300 psig range
 PCL-PM1KG    30,150.00 Kč   External pressure module, 0 to 1000 psig range
 PCL-PM3KG    30,150.00 Kč   External pressure module, 0 to 3000 psig range
 PCL-PM030A    30,150.00 Kč   External pressure module, 0 to 30 psia range
Obsah košíku Vyprázdnit košík Objednat
 * využijte možnost konzultovat ceny pøímo s našimi prodejci  (tel: 596 311 899 - prodejní oddìlení, e-mail: viz. kontaktní informace)
 Poznámka: Comes with test leads, protective rubber boot, NIST calibration certificate, 4 "AA" alkaline batteries, type K thermocouple and retractable thermocouple cable, and operators manual.

PCL-PM series pressure modules and PCL-PMA module adapters required for pressure measurement, sold separately.
 Pøíklad objednávky: (1) PCL-PM030G Pressure module, 0 to 30 psig, (1) PCL-PMA Pressure module adapter, (1) PCL1200 High-accuracy multifunction calibrator = 114,150.00 Kč

 Literatura
   Stáhnout kompletní specifikace produktu ve formátu PDF
 Manuály
   Stáhnout  PCL1200 - Multi-function Process calibrator
 Souvísející odkazy
   PCL-PM030G, PCL-PM3KG, PCL-PM015A: External Pressure Modules
For PCL340 Series, PCL1000 and PCL1200 Calibrators